Home > Specialty Glass > Information Technology > Optical Fiber Fluorescence Detector > Specifications and Options available
| Product Name | FLE1000![]() |
FLE500 (Available soon)![]() |
|---|---|---|
| Target fluorescent materials |
|
|
| Sensitivity | Detection Limit is arround 1nM (depending on the specimen and its shape) |
Detection Limit is arround 5nM (depending on the specimen and its shape) |
| Required amount of sample | 1 μL or less (even a DNA array chip of 0.1 mm in diameter can be measured) |
1μL or less |
| Data output | Analog or RS-232C output (Simple RS-232C datalogging software for WindousXP is available) |
Analog output |
| Response time | analog : 0.1sec(typ) RS-232C : 0.1sec |
0.2sec(typ) |
| Size | 130×160×40 mm | 100×75×30 mm |
| Power supply | DC24V (a converter from AC100 V is available) |
DC12V |
| Measurement spot | 0.2 mm in diameter (can be specified) |
0.2mm in diameter (can be specified) |
| Probe head | 4 mm in diameter (using the SELFOC® Micro Lens) |
4mm in diameter (using the SELFOC® Micro Lens) |
| Upper temperature limit of probe | Up to 110°C | Up to 110°C |
*This product can be customized as a fluorescence detection module for various types of analyzers.